Description
- It is an ecosystem that uses a Stirling cooler and does not require liquid nitrogen or freezing agents.
- No chemicals or refrigerants are required, significantly reducing running costs.
- When using 2ml vials, a maximum of 81 vials can be frozen. Cell bags can also be frozen by removing the freezing block.
- Lightweight and compact design, can be used inside a clean bench or isolator. It is also ideal for use in clean rooms and CPC/CPF.
- The cooling plate can be set without tools, which prevents temperature rise during sample transfer.
- The cooling area has a torque-hinged cover for easy opening and closing.
- The built-in heater allows the cooling plate to return to ambient temperature after the freezing process is complete.
- You can set and register your desired freezing pattern from the built-in touch panel.
- Plate/sample temperature can be measured and recorded, and processing results can be output in PDF and CSV format.
- Users can be managed with the user login function, and there are three levels of authority settings (FZ-2200 only).
- The optional remote access function allows you to check the processing status and download processing result data (PDF/CSV) from a web browser via LAN.
Note: This product cannot be exported to the United States, Europe, or South Korea. Please contact us for details.
Specification
Model | FZ-2100 | FZ-2200 |
---|---|---|
Max # of 2ml vials | 81 | |
Temp. control | PID control | |
Temp. sensor | Pt100 class B | |
LCD touch screen | Color 7″ (16:9) | |
Recovery heater | Max 20c | |
Resolution of cooling rate | -0.1c/min | |
Minimum temperature | -80℃ | |
Dimension of cooling area | W174 x D132 mm | |
Dimension | W480 x D340 x H460 mm | |
Power supply | AC100-240V 50/60Hz 150VA | |
Varidation (IQ/OQ) | N/A | Yes |
CFR21 Part11 compliant | N/A | Yes |
User management | N/A | 3 degree (Admin/Manager/User) |
History management | Yes | |
Report output | PDF/CSV |